YOU ARE VIEWING THE STAGING SITE

TEMPOS SH-3 dual-needle (3 cm)

TEMPOS SH-3 dual-needle (3 cm)

Measure the diffusivity and heat capacity of solid materials

The TEMPOS SH-3 needle is used to measure the thermal properties of solid materials. The TEMPOS low-power mode decreases the heat input to the needle, allowing you to measure frozen substances with a lower likelihood of phase change.

The SH-3 is ideal for measuring volumetric specific heat capacity and thermal diffusivity, and it performs well with high-viscosity liquids, moist soil, granular materials, and cured concrete or rock.

The SH-3 needle should be used with the included needle guide. Calibrate it using the SH-3 Delrin® verification block.

Technical specifications

Find specifications for the SH-3 dual-needle and other TEMPOS accessories on the TEMPOS page.

Thermal Conductivity
Range: 0.02–2.00 W/(m × K)
Accuracy: ±10% from 0.02–2.0 W/(m × K)

Thermal Resistivity
Range: 50–5000 °C

Thermal Diffusivity
Range: 0.1–1.0 mm2/s
Accuracy: ±10% at conductivity above 0.2 W/(m × K)
Accuracy: ±0.02 W/(m × K) from 0.10–0.20 W/(m × K)

Volumetric Specific Heat Capacity
Range: 0.5–4.2 MJ/m3
Accuracy: ±10% at conductivities above 0.1 W/(m × K)

Needle physical dimensions
Length: 30 mm
Diameter: 1.3 mm
Spacing: 6 mm

A photo of the TEMPOS thermal properties analyzer

Request a quote

Fill out the form below to help us pair you with the right expert. We’ll prepare the information you request, then contact you as soon as possible.